NewLab 100
Cloud Point
ASTM D5771
DIN 51597
EN 23015
EN 590
IP 444
Correlated:
ASTM D2500 - ASTM D5772 - ASTM D5773
IP 219 - IP 445 - IP 446
ISO 3015
JIS K2269
Cloud Point of Petroleum Products and Liquid Fuels.
This test method can determine the temperature of the test specimen at which wax crystals have formed sufficiently to be observed as a cloud with a resolution of 0.1 °C.
Measuring Cloud Point Principle
The sample is cooled down according to the methods while the clouds appearance is observed on the silver bottom of the test jar by means of an optical sensor. The measurement is done by reflection on the silver bottom of the test jar via a fast light detector. The signal from light detector is traded by the LabLink software. The dynamic measurement is performed regardless of the sample’s colour.
Measuring Cloud Point Devices
Light pulsed emission on I.R spectrum through a coaxial fibber optic.
Measuring Temperature Probe
- Platinum resistance PT100 class A.
- The PT100 is touching the bottom of the test jar.
Measuring Parameters
- Temperatures: in °C.
- Measuring range: +80°C … -80°C.
- Resolution: 0.06 °C.
- Accuracy: ± 0.1 °C.
- Repeatability / Reproducibility: as per standards methods or better.
Software Features
- New LabLink software able to manage up to 6 analytical heads simultaneously (stand alone).
- User friendly interface.
- All analytical parameters recorded.
- Customizable analysis parameters and methods.
- Customizable results report.
- Printable graphs and results.
- Self-identification of the typology of the analysers connected.
The software includes:
Analysis Menu
- Standard method as per ASTM / IP / ISO / EN / DIN… norms of reference.
- Optional methods:
- fast bath (to reduce the time of analysis)
- T-sample – T-bath (Delta T constant)
- cooling rate °C / h
- Audible alarm and displayed messages (at the end of the analysis and in case of errors and/or malfunctions).
Diagnostic Menu
- Direct access to all analog, digital, inputs and outputs.
- Selectable value displaying: °C / Volt.
Calibration Menu
- Automatic calibration of each temperature probe.
- Last calibration date referred to each single probe displayed and relative data printable.
- Display of calibration diagram.
- Insertion of offset values.
- Standard and advanced calibration modes.
Data Utilities
- Fields for introduction of operator and product name.
- Archive viewer for files recall.
- All analysis stored in Excel® compatible format.
- Storage capacity for more than 60’000 analysis.
- LIMS compatible.
Integrated Touch Screen Panel PC
- TFT/LCD 12".
- Resolution 1024 × 768, 16.2 M colours.
- 2 USB ports for connection to an external printer and/or external PC.
- Storage capacity for more than 60'000 analysis.
Test Jar
- Same dimensions and volume as described by the standard test methods.
- Product level mark.
- Small edge on the top in order to fix the glass cell to the analytical head.
- Silvered bottom with anti-scratch film protection.
Cooling System
- Integrated gas CFC free motor compressors:
- Single stage (for temperatures up to -40°C / 1)
- Double stage (for temperatures up to -80°C / 2)
- Equipped with an automatic energy power save system. After 15 minutes from the end of the analysis the cooling system goes in stand-by mode.
Safety Devices
- Pressure controller for 1st stage motor compressor.
- Pressure controller for 2nd stage motor compressor.
- Thermostat for 2nd stage activation.
- Thermo-switch for each cooling / heating jacket.
- Motor compressors equipped with internal overload devices.
Electrical Supply
- 220V ± 15% / 50 to 60 Hz.
- 115V ± 15% / 60 Hz.
Cord Cable
- 3 conductors flexible cable 2 m (7 feet) length with PVC sheath oil and heat resistant as per CENELEC directives.
Ambient Temperature
Spare Parts NewLab 100 SA
Suggested for 2 years operation.
- 5399: Fiber optic.
- 3399: PT100 product with connector Cloud Point.
- 1009: Test jar with silver bottom.
- 7315: O-ring for test jar.
Tools Required for Routine Calibration
Not needed for installation.
- 3013: Calibration decade box – PT100 simulator.
- 3102: Set of connectors and cables.
| Dimensions (cm) |
1 test position |
2 test positions |
3 test positions |
4 test positions |
6 test positions |
| width |
66 |
66 |
100 |
134 |
130 |
| depth |
60 |
60 |
60 |
60 |
75 |
| height |
80 |
80 |
80 |
80 |
170 |
| Weight (kg) |
| single stage compressor |
70 |
90 |
- |
- |
- |
| double stage compressor |
80 |
100 |
130 |
160 |
280 |
NewLab 100 ST
- Cooling system ST.
- Integrated gas CFC free Stirling motor compressor.
- For temperatures up to -110°C.
- Integrated Touch Screen Panel PC for ST.
- TFT/LCD 8”.
- Resolution 1024 × 768.
- Measuring range: +55°C … -95°C.
- Resolution: 0.01 °C
- Width: 34 cm.
- Depth: 60 cm.
- Height: 80 cm.
- Weight: 34 kg.
- Electrical supply :100∼240 Vac 50/60 Hz.
Spare Parts NewLab 100 ST
Suggested for 2 years operation.
- 5399: Fiber optic.
- 3399: PT100 product with connector Cloud Point.
- 1009: Test jar with silver bottom.
- 7315: O-ring for test jar.
Tools Required for Routine Calibration
Not needed for installation.
- 3013: Calibration decade box – PT100 simulator.
- 3102: Set of connectors and cables.
Linetronic Technologies SA
Via Onorio Longhi 2 • 6864 Arzo • Mendrisio • Switzerland
tel. +41 91 630 07 03 • fax +41 91 630 07 19
email: info@lin-tech.ch