Laboratory and Process Analyzers
Automatic Analysers → NewLab Range • OilLab Range • Thermofat • OnLine Range
NewLab X • Cloud Point • CFPP • Pour Point • Freezing Point (410) • Freezing Point (411) • Cloud and Pour Point • Filter Blocking Tendency • LTFT • Low-temperature Torque
NewLab 411
Freezing Point
ASTM D1177
Freezing Point of aviation fuels, aviation gasoline, aviation turbine fuels, engine coolants, antifreeze products, brake fluids, …
Solidification Point of Benzene.
Solidification Point of Industrial Organic Chemicals.
Measuring Freezing Point Principle
According to the methods, the sample is cooled down and stirred. The solid hydrocarbon crystals formation are detected by means of a light beam throught fiber optic reflected thanks to a mirror. As soon as crystals are detected, the sample is warmed up until their complete disappearance.
Measuring Freezing Point Devices
- Light pulsed emission on I.R spectrum through a coaxial fibber optic
- Coaxial fibber optic equipped with a mirror
Measuring Temperature Probe
- Platinum resistance PT100 class A
Stirrer
- A micro-motor drives all the mechanical system
- 3 coils stirrer made of brass
Measuring Parameters
- Temperatures: in °C / °F
- Measuring range: +55°C … +100°C
- Resolution: 0.01 °C
- Accuracy: ± 0.1 °C
- Repeatability / Reproducibility: as per standards methods or better
Software Features
New LabLink software able to manage up to 6 analytical heads simultaneously (stand alone)
- User friendly interface
- All analytical parameters recorded
- Customizable analysis parameters and methods
- Results report
- Printable graphs and results - any Windows® compatible printer can be used
The software includes:
Analysis Menu
- Standard method as per ASTM / IP / ISO / EN / DIN… norms of reference
- Optional methods:
- special detection of contaminants
- Audible alarm and displayed messages at the end of the analysis and in case of errors and/or malfunctions
Diagnostic Menu
- Direct access to all analog, digital, inputs and outputs
- Selectable value displaying: °C / Volt
Calibration Menu
- Automatic calibration of each temperature probe
- Last calibration date referred to each single probe displayed and relative data printable
- Display of calibration diagram
- Insertion of offset values
- Standard and advanced calibration modes
Data Utilities
- Fields for introduction of operator and product name
- Archive viewer for files recall
- All analysis stored in Excel® compatible format
- Storage capacity for more than 60’000 analysis
- LIMS compatible
Integrated Touch Screen Panel PC
- TFT/LCD 12"
- Resolution 1024 × 768, 16.2 M colours
- 2 USB ports for connection to an external printer and/or external PC
- Storage capacity for more than 60'000 analysis
Test Jar
- Insulated cooling jackets
- Integrated gas CFC/HCFC free liquid helium motor compressor
- Equipped with an automatic energy power save system
Cooling System
Integrated gas CFC free motor compressors:
- Single stage (for temperatures up to -40°C / 1)
- Double stage (for temperatures up to -80°C / 2)
Equipped with an automatic energy power save system. After 15 minutes from the end of the analysis the cooling system goes in stand-by mode.
Safety Devices
- Pressure controller
- Thermostat and thermo-switch
Electrical Supply
- 220V ± 15% / 50 to 60 Hz
- 115V ± 15% / 60 Hz
Cord Cable
- 3 conductors flexible cable 2 m (7 feet) length with PVC sheath oil and heat resistant as per CENELEC directives
Ambient Temperature
Dimensions and weight
- Width: 34 cm
- Depth: 60 cm
- Height: 80 cm
- Weight: 34 kg
Spare Parts
- 1210: Double tube 200 ml.
- 5235: Wire stirrer.
- 3596: PT100 product with connector.
Calibration Tools
- 3013: Calibration decade box – PT100 simulator.
- 3102: Kit of connectors and cables.
Linetronic Technologies SA
Via Onorio Longhi 2 • 6864 Arzo • Mendrisio • Switzerland
tel. +41 91 630 07 03 • fax +41 91 630 07 19
email: info@lin-tech.ch