Laboratory and Process Analyzers
Automatic Analysers → NewLab Range • OilLab Range • Thermofat • OnLine Range
Filterability of Lubricating Oils • Ring and Ball • Foaming Tester • Oxidation Stability • Evaporation Bath • RPVOT-570 • RPVOT-571 • RSSOT • Noack • Pensky Martens • Leonardo Pensky Martens • Golleo Abel+Pensky • Aniline Point • Penetration • R.E.C.C. •Abel • Abel-6500 • Cleveland • Cleveland-6700 • Tag • Tag-6900 • Reid Vapour Pressure • Vapour Pressure of Fuel • Ductilometer • Herschel Emulsifying • Saybolt Viscometer • Distillation
OilLab 6900 - OilLab 6901
Tag
ASTM D56
ASTM D3278
ASTM D3941
Subject
Flash point on petroleum products having a flash point between ambient temperature and +93°C. Suitable for flash point detection on different substances and (NO) waste materials, solvents…
OilLab 6900 / OilLab 6901 - Double / Single Automatic Tag Flash Point Analyser
- Two calibrated brass crucible with sample level mark (OilLab 6900).
- One calibrated brass crucible with sample level mark (OilLab 6901).
- Ignition system: gas or electric lighter.
- Control system: on microprocessor based.
- Temperature measurement: platinum resistance, PT100, Class A with SS sheath.
- Built-in barometric sensor with automatic barometric correction of results executed by the software.
- A single multi-detector combines the ionization detector and thermal detector.
- Heating: electrical heater with heating rates as per method.
Temperature range
- OilLab 6900: up to +110°C, lower temperature depending on the source connected to the instrument.
- OilLab 6901 +8°C … +110 °C.
Cooling
- OilLab 6900: External cooling joints, mandatory for granting temperature below ambient.
- OilLab 6901: via internal Peltier Module.
Safety features
- Automatic diagnostic in case of breakage of the sample temperature probe and thermal sensors.
- Automatic fire detection and extinguisher system.
- Overheating protection with auto shut off during the test.
- Gas valve for closing the gas supply.
Built-in Touch Screen Panel PC
- For the managing of the analyser by means of the Lab-Link software.
- 10.1” Fanless Panel Pc Intel® quad-core processor.
- PCAP touch screen: finger, gloves and tft stylus compatible.
- Resolution 1024 × 768 and 16.2 M colours.
- 2 × USB for connection to an external printer.
- 1 × Ethernet port for LAN or Lims network.
- Storage capacity for more than 65’000 analysis.
Lab-Link operating software
- Automatic execution of the analysis as per method.
- Automatic handling of samples with unknown flash point.
- Display in real time of all the analysis parameters and status.
- Field for introduction of operator and product name.
- Expected flash point temperature programmable.
- Audible alarm and displayed message for end analysis, errors/malfunctions.
- Configuration menu with up to 20 preset sample and expected flash point.
- Automatic barometric correction of result.
- USB flash memory pen drive with the installed software with calibration and settings (copy of backup), User Manual (PDF format).
Electrical Supply
- Cord cable with Schuko plug.
- 220 V / 50 Hz.
- 115 V / 60 Hz.
Ambient Temperature
Dimensions
- width 49 cm.
- depth 60 cm.
- height 32 cm.
Weight
Calibration Tools
- Testo511: absolute barometer.
- 3013: calibration decade box - PT100 Simulator.
- 17587: kit of connectors and cables for OilLab 6000 series.
Spare Parts
- 5755: calibrated brass crucible with handle.
- 17314: PT100 product.
- 3761: electrical ignitor long life.
- 3646: thermal ionisation PT100 sensor.
Accessories
- LT/CB-40830/M-PLUS: Laboratory cryostat bath for temperatures up to -30°C.
- Pump for external circulating option.
- Solid steel structure coated in epoxy anti-acid paint with upper side made in stainless-steel.
- Internal chamber in seamless stainless steel with rounded corners for efficient circulation.
- Easy cleaning thanks to atmospheric drain on the bottom of the chamber.
- Control head equipped with digital display showing set and actual temperature, display resolution 0,1°C.
- Temperature range from -30°C to +99,9°C accuracy to ±0,5°C to +37°C.
- Manual safety thermostat against overtemperature.
- Circulating pump: prevalence 4m / 2 l - min ~ prevelance 1 m /12 l - min.
- Weight: 20 kg (empty).
- Dimensions: W 335 × D 470 × H 630 mm.
- Capacity 8 liters.
- Power supply: 220 Vac or 115 Vac 50/60 Hz.
- Max. consumption 1500 Watt.
- DIN 12880 class 2 construction type.
Reference Sample
- 7271: Certified reference material Tag closed cup.
- Product nature: Jet A1.
- Certified value: + 40.0 °C approx.
- Pack size: 250 ml.
Linetronic Technologies SA
Via Onorio Longhi 2 • 6864 Arzo • Mendrisio • Switzerland
tel. +41 91 630 07 03 • fax +41 91 630 07 19