Filterability of Lubricating Oils • Ring and Ball • Foaming Tester • Oxidation Stability • Evaporation Bath • RBOT & TFOUT • RPVOT • Noack • Pensky Martens • Leonardo Pensky Martens • Golleo Abel+Pensky • Aniline Point • Penetration • R.E.C.C. • Abel • Cleveland • Tag • Tag - 6901 • Reid Vapour Pressure • Ductilometer • Herschel Emulsifying • Saybolt Viscometer • Distillation
OilLab 6901 - Tag
ASTM D56
ASTM D3278
ASTM D3941
Subject
Flash point on petroleum products having a flash point between ambient temperature and +93°C. Suitable for flash point detection on different substances and (NO) waste materials, solvents…
Automatic Tag Flash Point Analyser Stand Alone
- Calibrated brass crucible with sample level mark.
- Ignition system: gas or electric lighter.
- Control system: on microprocessor based.
- Temperature measurement: platinum resistance, PT100, Class A with SS sheath.
- Built-in barometric sensor with automatic barometric correction of results executed by the software.
- A single multi-detector combines the ionization detector and thermal detector.
- Heating: electrical heater with heating rates as per method.
- Temperature range: +8°C … +110 °C.
- Cooling: inbuilt forced air fan for automatic cooling at the end of test.
- Safety features:
- Automatic diagnostic in case of breakage of the sample temperature probe and thermal sensors;
- Automatic fire detection and extinguisher system;
- Overheating protection with auto shut off during the test;
- Gas valve for closing the gas supply.
- Built-in Touch Screen Panel PC for the managing of the analyser by means of the Lab-Link software:
- TFT/LCD 8.4”;
- Resolution 1024 × 768 and 16.2 M colours;
- 2 × USB for connection to an external printer;
- 1 × Ethernet port for LAN or Lims network;
- Storage capacity for more than 65’000 analysis.
- Lab-Link operating software:
- Automatic execution of the analysis as per method;
- Automatic handling of samples with unknown flash point;
- Display in real time of all the analysis parameters and status;
- Field for introduction of operator and product name;
- Expected flash point temperature programmable;
- Audible alarm and displayed message for end analysis, errors/malfunctions;
- Configuration menu with up to 20 preset sample and expected flash point;
- Automatic barometric correction of result.
- Cord cable with Schuko plug.
- USB flash memory pen drive with the installed software with calibration and settings (copy of backup), User Manual (PDF format).
Electrical Supply
- 220V ± 15% / 50 Hz
- 115V ± 15% / 60 Hz
Ambient Temperature
Dimensions
- width 37 cm
- depth 52 cm
- height 32 cm
Weight
Accessories
- 5050: gas reducer 30 mbar.
- 5052: gas connection tube, 5 m.
Calibration Tools
- Testo511: absolute barometer.
- 3013: calibration decade box - PT100 Simulator.
- OilLab 91: kit of connectors and cables.
Spare Parts
- LAB-6900/09-05: calibrated brass crucible with handle.
- LAB-6900/08-12: PT100 product.
- LAB-6900/07-01: electrical ignitor long life.