Laboratory and Process Analyzers
Automatic Analysers → NewLab Range • OilLab Range • Thermofat • OnLine Range
Filterability of Lubricating Oils • Ring and Ball • Foaming Tester • Oxidation Stability • Evaporation Bath • RPVOT (570) • RPVOT (571) • Noack • Air Release • Pensky Martens • Leonardo Pensky Martens • Golleo Abel+Pensky • Aniline Point • Penetration • R.E.C.C. • Abel • Cleveland • Tag • Tag - 6901 • Reid Vapour Pressure • Ductilometer • Herschel Emulsifying • Saybolt Viscometer • Distillation
OilLab 650 - OilLab 650 Plus
Abel
EN 924 - EN 13736
IP 170 - IP 491 - IP 492
ISO 1516 - ISO 3679 - ISO 13736
Subject
Flash point on petroleum products having a flash point between -18°C and 71°C (kerosene and solvents). Suitable for flash point detection on different substances and waste materials, solvents…
Measuring Abel Principle
The sample is warmed up according to the methods. When the sample reaches the selected test temperature, the shutter is opened and the ignition system introduces itself automatically. If the flash point is reached, the detection is done by an ionisation detector. If not, the shutter closes again and the sample continues to warm up until the next test temperature.
Measuring Abel Devices
- Measurement of the Flash Point detected by an ionisation detector.
- Testing unit equipped with two ignition systems.
- Electrical system or flame exposure device.
Measuring Temperature Probe
- Platinum resistance PT100 class A.
- Temperatures: in °C.
- Measuring range: -50°C … +120°C.
- Resolution: 0.06 °C.
- Accuracy: ± 0.1 °C.
- Repeatability / Reproducibility: as per standards methods or better.
Measuring Parameters
- Measuring range: -50°C … +120°C (PT100 sensor capacity).
- Instrument range: +10°C … +75°C (as per method).
Software Features
- User friendly interface.
- All analytical parameters recorded.
- Customizable analysis parameters and methods.
- Customizable results report.
- Printable graphs and results.
- Self-identification of the typology of the analysers connected.
- Warning if results obtained is out of specification.
The software includes:
Analysis Menu
- Standard method as per ASTM / IP / ISO / EN / DIN… norms of reference.
- Unknow sample.
- Audible alarm and displayed messages at the end of the analysis and in case of errors and/or malfunctions.
Diagnostic Menu
- Direct access to all analog, digital, inputs and outputs.
- Selectable value displaying: °C / Volt.
Calibration Menu
- Automatic calibration of each temperature probe.
- Last calibration date referred to each single probe displayed and relative data printable.
- Display of calibration diagram.
- Insertion of offset values.
- Standard and advanced calibration modes.
Data Utilities
- Fields for introduction of operator and product name.
- Archive viewer for files recall.
- All analysis stored in Excel® compatible format.
- Storage capacity for more than 60’000 analysis.
- LIMS compatible.
Integrated Touch Screen Panel PC
- TFT/LCD 8".
- Resolution 1024 × 768, 16.2 M colours.
- USB port for connection to an external printer and/or external PC.
- Storage capacity for more than 60'000 analysis.
Test Cup
- The cup is made of brass provided with high temperature resistant handle.
- Sample level mark.
Heating
- Electrical heater.
- Equipped with over temperature cut-out.
Cooling System
- Liquid refrigerant controlled by internal solenoid valve (OilLab 650), external cooling circulator or tap water is required.
- Peltier elements granting a temperature to +10°C (OilLab 650 Plus).
Stirrer
- An electric motor drives a flexible transmission coil allowing the stirring of the product.
Shutter
- Automatic mechanism opening the shutter conform to the methods.
Safety Features:
- Gas valve for closing the gas supply (max 30 mBar), at the end of the test.
- Overheating protection with auto shut off during the test.
- Auto fire detection by means of thermal fuse with acoustic alarm.
Electrical Supply
- 220V ± 15% / 50 to 60 Hz
- 115V ± 15% / 60 Hz
Cord Cable
- 3 conductors flexible cable with schuko plug
Ambient Temperature
Dimensions
- width 37 cm
- depth 48 cm
- height 61 cm
Weight
Accessories
- 5050: Gas reducer 30 mbar.
- 5052: Gas connection tube, 5 m.
- LT/CB-40800/M-30: Laboratory Cryostat Bath.
- For temperatures up to -30°C.
- Solid steel structure coated in epoxy anti-acid paint with upper side made in stainless-steel.
- Internal chamber in seamless stainless steel with rounded corners for efficient circulation.
- Easy cleaning thanks to atmospheric drain on the bottom of the chamber.
- Control head equipped with digital display showing set and actual temperature, display resolution 0,1°C.
- Temperature range from -30°C to +99,9°C, accuracy to ±0,5°C to +37°C.
- Manual safety thermostat against overtemperature.
- Circulating pump: prevalence 4mt/2lt.-min ~ prevalence 1mt/12lt.-min.
- Weight: 20 kg (empty).
- Dimensions: width 33.5 cm, depth 47.0 cm, height 63.0 cm.
- Internal bath dimensions: width 10.5 cm, depth 12.5 cm, height 15.0 cm.
- Capacity: 8 liters.
- Power supply: 220 Vac or 115 Vac 50/60 Hz.
- Max. consumption: 1500 Watt.
- DIN 12880 class 2 construction type.
Spare Parts
- 7175: PT100 Sensor, A class, made in stainless steel with connector.
- 17028: Security thermal fuses.
- 5666: Calibrated brass crucible.
- 7369: Detection - ionisation cable.
- 3754: Thermal detector.
- 7172: Electrical ignitor long life.
- 7170: Flexible stirrer with connection joints.
Calibration Tools
- 3013: calibration decade box – PT100 simulator.
- 3102: set of connectors and cables for cold range.
Linetronic Technologies SA
Via Onorio Longhi 2 • 6864 Arzo • Mendrisio • Switzerland